6515-01-659-6866

ANALYZER,INTRAVENOUS INFUSION PU


6515-01-659-6866 (6515016596866, 016596866)
Part Number (1 listing): IDA-5/2SHK250V
Summary

FSG 65 / FSC 6515

DEMIL: A

DEMIL INT: 1

HMIC: N

Date Created: 9 Dec 2016

Request a Quote


;



MASTER CROSS REFERENCE LIST (MCRL) | NSN 6515-01-659-6866 Submit RFQ
Manufacturer Part NumberRNCCRNVCManufacturer NameCAGE
IDA-5/2 SHK250V3 - Design Control Reference2 - Production ItemFLUKE BIOMEDICAL CORPORATION  34511      
Code Definitions

3 - Design Control Reference

Reference Number Category Code (DRN_2910)


Design Control Reference. The primary number used to identify an item of production or a range of items of production, by the manufacturer (individual, company, firm, corporation, or Government activity) which controls the design, characteristics, and production of the item by means of its engineering drawings, specifications and inspection requirements.

2 - Production Item

Reference Number Variation Code (DRN_4780)


A design control or other reference number that is an item-identifying number for an item of production, or a source control reference or a specification or standard part, type, or similar reference number that is an item-identifying number for an item of supply.


FLIS - SEG A - IDENTIFICATION | NSN 6515-01-659-6866 Submit RFQ
FSGFSCNIINFIIGINCCRITICALITYTYPE OF ITEMMRCDEMILHMICEDCPMICADPECDATE CREATED
656515016596866T133-F47433X43ANAA09 Dec 2016
Code Definitions

65

Federal Supply Group (FSG_3994/FSC_WI_FSG_3996)


Medical, Dental, and Veterinary Equipment and Supplies

6515

Federal Supply Class (FSC_3994/FSC_WI_FSC_3996)


Medical and Surgical Instruments, Equipment, and Supplies

016596866

National Item Identification Number (NCB_CD_4130/I_I_NBR_4131)


NSN 6515-01-659-6866

T133-F

Federal Item Identification Guide (FIIG_4065)

47433

Item Naming Code (INC_4080)

9 Dec 2016

NIIN Assignment Date (DT_NIIN_ASGMT_2180)

X

Item Criticality (CRITL_CD_FIIG_3843)


The item does not have a nuclear hardened feature or any other critical feature such as tolerance, fit restriction or application.

4

Type of Item Identification (TYP_II_4820)


Partial Descriptive Item Identification (Type 1 Concept)

3

Reference / Partial Descriptive Method Reason Code (RPDMRC_4765)


An approved item name and FIIG exist but the item involved is so unique in design that it cannot be fully described in accordance with the FIIG.

A

Demilitarization Code (DEMIL_CD_0167)


Non-MLI/ Non-SLI — Demilitarization not required.

N

Hazardous Material Indicator Code (HMIC_0865)


Indicates there is no data in the HMIRS and the NSN is in a FSC not generally suspected of containing hazardous materials.

A

Electrostatic Discharge Code (ESD_EMI_CD_2043)


No known Electrostatic Discharge (ESD) or Electromagnetic Interference (EMI) sensitivity.

A

Precious Metals Indicator Code (PMIC_0802)


Item does not contain Precious Metal.

0

ADPE Identification Code (ADPEC_0801)


Represents items with no ADP components. NOTE: Codes 1 through 6 are only to be used when the item is Automatic Data Processing Equipment (ADPE) in its entirety and is limited to the type meeting only one of the definitions for codes 1 through 6. (See code 9)


FLIS - SEG B - MOE RULE DATA | NSN 6515-01-659-6866 Submit RFQ
MOE RuleFormer MOEAMCAMSCNIMSCIMCIMC ActivityAACDSORCollaboratorsReceiversEffective Date
ZW011 Dec 2016
Code Definitions

FLIS - SEG G - FREIGHT DATA | NSN 6515-01-659-6866 Submit RFQ
IntegrityOriginatingHMCLCLWCCTCCSHCADCACCASHLTLNMFCNMFC Sub ItemUFCFreight DescriptionRail Variance
ZW
Code Definitions

ZW

Originating Activity Code (ORIG_ACTY_CD_4210)


Turkey

Rail Variance

Rail Variance (RAIL_VARI_CD_4760)


No Variation. NMFC and UFCC Match.


FLIS - SEG M/V - CHARACTERISTICS DATA | NSN 6515-01-659-6866 Submit RFQ
MRCDecoded RequirementClear Text Reply
ACDCCurrent TypeAC
FREQFrequency In Hertz50.0 AND 60.0
CXCYPart Name Assigned By Controlling AgencyINFUSION DEVICE ANALYZER
ABKWOverall Height12.0 INCHES NOMINAL
ABMKOverall Width8.0 INCHES NOMINAL
FEATSpecial FeaturesTESTS UP TO 4 INFUSION PUMPS AT THE SAME TIME;COMPATIBLE WITH VIRTUALLY ANY TYPE OF INFUSION DEVICE;BUILT IN MEMORY SAVES TEST RESULTS;SINGLE FLOW DUAL FLOW PIGGYBACK AND PCA TESTING;INSTANTANEOUS AND AVERAGE FLOW MEASUREMENT OF UP TO 1500 ML PER HR;OCCLUSION PRESSURE MEASUREMENTS TO 45 PSI;OPERATING 15 DEGREE C TO 30 DEGREE C;STORAGE NEGATIVE 20 DEGREE C TO PLUS 40 DEGREE C WHEN DRAINED OF ALL LIQUID;HUMIDITY 10 PERCENT TO 90 PERCENT NON CONDENSING
ELECVoltage In Volts100.0 AND 240.0


Back to Top