TM 9-6920-430-14
(45) Slowly open the gas supply bottle valve (41) two turns counterclockwise.
(46) Check to assure that OUTLET BLEED valve (42) and INLET BLEED
valve (48) are closed.
(47) Open the OUTLET SHUTOFF valve (43) two turns counterclockwise.
(48) Open the INLET SHUTOFF valve (44) two turns counterclockwise. The
INLET PRESSURE (45) and OUTLET PRESSURE (46) gages and the digital readout
device (47) should all read approximately the same ± 200 psi.
(49) Check the following high pressure fittings for leaks using leak detecting
fluid. If leaks are detected, do steps 50 through 54. If no leaks are detected, go to step 55.